Segments:measure:table:select, Segments:measure:table:list – GW Instek GDS-2000A series Programming Manual User Manual
Page 245

COMMAND DETAILS
245
Related
commands
:SEGMents:MEASure:MODe ; :SEGMents:MEASure:T
ABle:SELect ; :SEGMents:MEASure:TABle:LIST
Parameter/
Return parameter
CH1~CH4
Channel 1 to 4
Example1
:SEGMents:MEASure:TABle:SOURce CH1
Sets the source to CH1.
:SEGMents:MEASure:TABle:SELect
Description
Sets or queries a segment to view in the
measurement table.
Syntax
:SEGMents:MEASure:TABle:SELect {<NR1> | ? |}
Related
commands
:SEGMents:TOTalnum
Parameter
<NR1>
1~2048
Return parameter Returns the number of segments as <NR3>.
Example1
:SEGMents:MEASure:TABle:SELect 10
Select segment number 10.
:SEGMents:MEASure:TABle:LIST
Description
Returns the measurement results of each segment
in the list.
Syntax
:SEGMents:MEASure:TABle:LIST?
Return parameter
Returns the measurements results as a string for
each segment.
Example
:SEGMents:MEASure:TABle:LIST?
>"GW GDS-2064A, serial number P930116, version
V1.11",Segment Summary : CH1, Seg.,Pk-Pk
(V),Pk-Pk (V),1,8.00m,8.00m……etc
Set
Query