Trigger:runt:when, Trigger:runt:time – GW Instek GDS-2000A series Programming Manual User Manual
Page 128

GDS-2000A Programming Manual
128
NEGative
Negative polarity
EITher
Positive or negative polarity
Return parameter
Returns the pulse runt trigger polarity.
Example
:TRIGger:RUNT:POLarity POSitive
Sets the Pulse Runt trigger polarity to positive.
:TRIGger:RUNT:WHEn
Description
Sets or queries the Pulse Runt trigger conditions.
Syntax
:TRIGger:RUNT:WHEn {MOREthan| LESSthan | EQual
| UNEQual | ? }
Related
commands
:TRIGger:TYPe
:TRIGger:RUNT:TIMe
Parameter
MOREthan
>
LESSthan
<
Equal
=
UNEQual
≠
Return parameter
Returns the pulse runt trigger condition.
Example
:TRIGger:RUNT:WHEn UNEQual
Sets the Pulse Runt trigger condition to unequal
(≠).
:TRIGger:RUNT:TIMe
Description
Sets or queries the Pulse Runt trigger time.
Syntax
:TRIGger:RUNT:TIMe {<NRf> | ? }
Related
commands
:TRIGger:TYPe
:TRIGger:RUNT:WHEn
Parameter
<NRf>
Pulse runt time (4nS to 10S)
Return Parameter <NR3>
Returns the runt time in seconds.
Set
Query
Set
Query