Measurement Performance

During meter verification, you can choose between setting digital communications process variable
values to the configured fault indicator or the last measured value. The values will remain fixed for
the duration of the test (approximately four minutes). Disable all control loops for the duration of the
procedure, and ensure that any data reported during this period is handled appropriately.

To perform meter verification:

Using ProLink II, follow the procedure illustrated in Figure 10-1.

Using the display menu, follow the procedure illustrated in Figure 10-2. For a complete
illustration of the meter verification display menu, see Figure B-8.

Using a DeviceNet tool, follow the procedure illustrated in Figure 10-3.

For a discussion of meter verification results, see Section 10.3.1.

Figure 10-1

Meter verification procedure – ProLink II

Verify configuration

parameters

Tools >
Meter Verification >
Structural Integrity Method

View previous test data

Next

Enter optional test data

Initialize and start meter

verification

Next

Abort

Next

Fault

configuration

Hold last

value

Progress bar shows

test in progress

Next

Finish

(2)

Graph of results

Rerun

test?

Yes

Pass

Fail

No

Start

Back

(1)

View report (option to print

or save)

Back

Abort

(1) If the graph was viewed at the beginning of the procedure,

clicking Back here will return to the beginning of the procedure
(along the dotted line).

(2) The results of the meter verification test are not saved until

Finish is clicked.