4 obtaining external pressure and temperature data, Obtaining external pressure and temperature data, Section 9.4 – Emerson Process Management MICRO MOTION 2400S User Manual

Page 95: See section 9.4, Vicenet tool, see figure 9-4

Advertising
background image

Configuration and Use Manual

87

Pressure Compensation and Temperature Compensation

Me

asu

reme

nt P

er
form

anc

e

Defaults

T

rou
b

lesho

o

ti

ng

Compensation

Figure 9-4

Configuring external temperature compensation with a DeviceNet tool

9.4

Obtaining external pressure and temperature data

The DeviceNet output assemblies used to obtain external pressure and/or temperature data are listed
in Table 9-1. Use standard DeviceNet methods to implement the required connection.

Table 9-1

Output assemblies used for pressure or temperature compensation

Instance ID

Data description

Size

Bytes

Data type

50

External pressure

4 bytes

Bytes 0–3

REAL

51

External temperature

4 bytes

Bytes 0–3

REAL

52

External pressure
External temperature

8 bytes

Bytes 0–3
Bytes 4–7

REAL
REAL

Enable temperature

compensation

Class: Calibration Object (0x65)
Instance: 1
Attribute ID: 25
Data type: BOOL
Value:
· 0 = disabled
· 1 = enabled

Set temperature unit

Class: Analog Input Point Object (0x0A)
Instance: 4
Attribute ID: 102
Value: See Table 6-6
Service: Set

Yes

Use static

temp value?

No

Set up output

assembly

(2)

Class: Calibration Object (0x65)
Instance: 1
Attribute ID: 26
Data type: REAL
Service: Set

Set static value

(1) Temperature measurement unit

must be configured to match
temperature unit used by
external device. See
Section 6.3.

(2) See Section 9.4.