Emerson Process Management MICRO MOTION 2400S User Manual

Page 162

Advertising
background image

154

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Device Profile

32

9-wire cable
RTD resistance

REAL

Get

V

The resistance of the
9-wire cable

Unit = ohms

33

Meter RTD
resistance

REAL

Get

V

The resistance of the
meter RTD

Unit = ohms

34

Number of
power cycles

UINT

Get

V

The number of transmitter
power cycles

35

Power on time

Unsigned 32

Get
Reset

(1)

V

The cumlative amount of
time the tranmitter has
been on since the last
reset (Class 0x01,
Attribute 0x05)

Seconds since last
reset

36

Line RTD

REAL

Get

V

The resistance of the
process line RTD

Unit = ohms

37

Actual target
amplitude

REAL

Get

V

The amplitude the
transmitter is attempting
to drive the sensor

Unit = mV/HZ

38

Input voltage

REAL

Get

V

The number of volts on
the power input terminals

Unit = volts

39

Drive current

REAL

Get

V

The drive current

Unit = milliamps

40

Alarm 7

UINT

Get

V

A collection of status bits

• 0x0001 = K1/FCF

Combination
Unrecognized

• 0x0002 = Warming

Up

• 0x0004 = Low Power
• 0x0008 = Tube not

Full

• 0x0010 = Meter Ver

Fault

• 0x0020 = Meter Ver

Info

• 0x0040 =UI PROM

error

• 0x0080 = Not Used
• 0x0100 = Not Used

0x0200 = Not Used

• 0x0400 = Not Used
• 0x0800 = Not Used
• 0x1000 = Not Used
• 0x2000 = Not Used
• 0x4000 = Not Used
• 0x8000 = Not Used

Table C-7

Diagnostics Object (0x66) – Instance 1 continued

Attrib
ID

Name

Data type

Service

Mem

Description

Comments