2 using the network led, 5 viewing transmitter status, 1 using the status led – Emerson Process Management MICRO MOTION 2400S User Manual

Page 50: Using the network led, Viewing transmitter status 7.5.1, Using the status led, See section 7.5, D, see section 7.4.2, Led, see section 7.5.1, Table 7-3 lists the dif

Advertising
background image

42

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Using the Transmitter

7.4.2

Using the network LED

The behavior of the network LED is standard, and is defined by the DeviceNet protocol. Table 7-4
lists the different states of the network LED and defines each state.

7.5

Viewing transmitter status

You can view transmitter status using the status LED, ProLink II, or a DeviceNet tool. Depending on
the method chosen, different information is displayed.

7.5.1

Using the status LED

The status LED shows transmitter status as described in Table 7-5. Note that the status LED does not
report event status or alarm status for alarms with severity level set to Ignore (see Section 8.8).

Table 7-3

Module LED states, definitions, and recommendations

Module LED state

Definition

Recommendations

Off

No power

Check the connection to the DeviceNet network.

Solid green

No processor faults

No action required.

Flashing green

Needs DeviceNet configuration;
may be in Standby state

Indicates an A006 alarm. Characterization
parameters are missing. See Section 6.2.

Solid red

Non-recoverable fault

Power cycle the transmitter. If condition does not
clear, call Micro Motion customer service.

Flashing red

Recoverable fault

Check for any status alarms.

Flashing red/green

Device in self-test

Wait until self-test is complete.
Check the Identity Object (0x01) for device states.

Table 7-4

Network LED states, definitions, and recommendations

Network LED state

Definition

Recommendations

Off

Device not online

The device is not connected to the network.

(1)

Check

the wiring if this LED is lit.

(1) If the transmitter is the only device on the network, and there is no host on the network, this is the expected LED state, and no action

is required.

Solid green

Device online and connected

No action required.

Flashing green

Device online but not connected

The device is connected to the network, but has not
been allocated by a host. No action required.

Solid red

Critical link failure

The most common cause is duplicate MAC IDs
(node addresses) on the network. Check for
duplicate MAC IDs.
Other causes include incorrect baud rate setting or
other network failure.

Flashing red

Connection timeout

Power cycle the device, or release and re-allocate
the device from the DeviceNet master.
If desired, increase the timeout value (Expected
Packet Rate) in the DeviceNet Object (0x03).

Flashing red/green

Communication faulted state

Not implemented in the Model 2400S DN transmitter.

Advertising