Llustrated in figure 10-3 – Emerson Process Management MICRO MOTION 2400S User Manual

Page 102

Advertising
background image

94

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Measurement Performance

Figure 10-3

Meter verification procedure – DeviceNet tool

Step 2

Set uncertainty limit (optional)

Step 3

Start/abort procedure

Step 1

Set output state (optional)

Able to

complete?

Step 9

Check abort code

CAUTION

Yes (>0)

Step 4

Check current algorithm state

Running?

Step 5

Read percent complete

Yes (=16)

Step 7

Check inlet stiffness

Within limits?

No (>0)

Yes (=0)

Step 8

Check outlet stiffness

Within limits?

No (>0)

Yes (=0)

CAUTION

PASS

No (<16)

No (=0)

Manual abort (optional)

Step 6

Check algorithm abort state

See Table 10-1 for the device profile
information for each step.