1 obtaining the test point values, 2 evaluating the test points, Obtaining the test point values – Emerson Process Management MICRO MOTION 2400S User Manual

Page 131: Evaluating the test points

Advertising
background image

Configuration and Use Manual

123

Troubleshooting

Me

asu

reme

nt P

er
form

anc

e

Defaults

T

rou
b

lesho

o

ti

ng

Compensation

11.19.1

Obtaining the test point values

To obtain the test point values:

With the display, configure the required test points as display variables. See Section 8.9.5.

With ProLink II

a.

Click

ProLink

>

Diagnostic Information

.

b.

Observe or record the values displayed for

Tube Frequency

,

Left Pickoff

,

Right Pickoff

,

and

Drive Gain

.

With a DeviceNet tool, execute a Get for the attributes listed in Table 11-4.

11.19.2

Evaluating the test points

Use the following guidelines to evaluate the test points:

If the drive gain is erratic, negative, or saturated, refer to Section 11.19.3.

If the value for the left or right pickoff does not equal the appropriate value from Table 11-5,
based on the sensor flow tube frequency, refer to Section 11.19.4.

If the values for the left and right pickoffs equal the appropriate values from Table 11-5, based
on the sensor flow tube frequency, record your troubleshooting data and contact Micro Motion
customer service. See Section 11.3.

Table 11-4

Test points with DeviceNet tool

Test point

Class

Instance

Attribute

Drive gain

Diagnostics Object (0x66)

1

20

Tube period

21

Left pickoff

23

Right pickoff

24

Table 11-5

Sensor pickoff values

Sensor

(1)

(1) If your sensor is not listed, contact Micro Motion. See

Section 11.3

.

Pickoff value

ELITE

®

CMF sensors

3.4 mV peak-to-peak per Hz based on sensor flow tube frequency

F025, F050, F100 sensors

3.4 mV peak-to-peak per Hz based on sensor flow tube frequency

F200 sensors

2.0 mV peak-to-peak per Hz based on sensor flow tube frequency

H025, H050, H100 sensors

3.4 mV peak-to-peak per Hz based on sensor flow tube frequency

H200 sensors

2.0 mV peak-to-peak per Hz based on sensor flow tube frequency

R025, R050, or R100 sensors

3.4 mV peak-to-peak per Hz based on sensor flow tube frequency

R200 sensors

2.0 mV peak-to-peak per Hz based on sensor flow tube frequency

T-Series sensors

0.5 mV peak-to-peak per Hz based on sensor flow tube frequency

CMF400 I.S. sensors

2.7 mV peak-to-peak per Hz based on sensor flow tube frequency