In table 8-5 – Emerson Process Management MICRO MOTION 2400S User Manual

Page 70

Advertising
background image

62

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Optional Configuration

3. Assign a process variable to the event (Attribute 10).

4. Specify the event’s setpoint(s) – the value(s) at which the event will occur or switch state (ON

to OFF, or vice versa).

If Event Type is High or Low, only Setpoint A is used (Attribute 8)

If Event Type is In Range or Out of Range, both Setpoint A and Setpoint B (Attributes 9
and 10) are required.

5. Assign the event to an action or actions, if desired. Possible actions are listed in Table 8-6. To

do this:

Using ProLink II, open the Discrete Input panel in the Configuration window, identify the
action to be performed, then specify the event using the dropdown list. See Figure B-3.

Note: For consistency with other Micro Motion products, the Discrete Input panel is used here even
though the Model 2400S DN transmitter does not provide a discrete input.

Using the display, see Figure B-6 and use the ACT submenu.

Using a DeviceNet tool, refer to Table C-7, use Attribute 84 to specify the action to be
performed, and set Attribute 85 to specify which event will initiate the action.

Table 8-5

Event types

Type

DeviceNet
code

Description

High (> A)

0

Default. Discrete event will occur if the assigned variable is greater than the
setpoint (A).

(1)

(1) An event does not occur if the assigned variable is equal to the setpoint.

Low (< A)

1

Discrete event will occur if the assigned variable is less than the setpoint (A).

(1)

In Range

2

Discrete event will occur if the assigned variable is greater than or equal to the low
setpoint (A) and less than or equal to the high setpoint (B).

(2)

(2) An event occurs if the assigned variable is equal to the setpoint.

Out of Range

3

Discrete event will occur if the assigned variable is less than or equal to the low
setpoint (A) or greater than or equal to the high setpoint (B).

(2)

Table 8-6

Event actions

ProLink II label

Display label

DeviceNet
code

Description

Start sensor zero

START ZERO

1

Initiates a zero calibration procedure

Reset mass total

RESET MASS

2

Resets the value of the mass totalizer to 0

Reset volume total

RESET VOL

3

Resets the value of the liquid volume totalizer to 0

(1)

Reset gas std volume total

RESET GSV

21

Resets the value of the gas standard volume totalizer
to 0

(2)

Reset API ref vol total

RESET TCORR

4

Resets the value of the API temperature-corrected
volume totalizer to 0

(3)

Reset ED ref vol total

RESET STD V

5

Resets the value of the ED standard volume totalizer
to 0

(4)

Reset ED net mass total

RESET NET M

6

Resets the value of the ED net mass totalizer to 0

(4)

Reset ED net vol total

RESET NET V

7

Resets the value of the ED net volume totalizer to 0

(4)

Advertising