Emerson Process Management MICRO MOTION 2400S User Manual

Page 160

Advertising
background image

152

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Device Profile

15

Alarm status 4

UINT

Get

V

A collection of status bits

• 0x0001 = API:

Temperature out of
range

• 0x0002 = API:

Density out of range

• 0x0004 = Line RTD

out of range

• 0x0008 = Meter RTD

out of range

• 0x0010= Reverse

flow

• 0x0020 = Factory

data error

• 0x0040 = ED: bad

curve

• 0x0080 = LMV

override

• 0x0100 = ED:

Extrapolation error

• 0x0200 = Need

calibration factor

• 0x0400 = Non-volatile

memory error (2700)

• 0x0800 = RAM error

(2700)

• 0x1000= Transmitter

not characterized

• 0x2000 = Non-volatile

memory error (CP)

• 0x4000 = Non-volatile

memory error (CP)

• 0x8000 = Non-volatile

memory error (CP)

16

Alarm status 5

UINT

Get

V

A collection of status bits

• 0x0001 = Boot sector

(CP)

• 0x0002 = Not used
• 0x0004 = Not used
• 0x0008 = Not used
• 0x0010 = Not used
• 0x0020 = Not used
• 0x0040 = Not used
• 0x0080 = Not used
• 0x0100 = D3

calibration in progress

• 0x0200 = D4

calibration in progress

• 0x0400 =

Temperature slope
calibration in progress

• 0x0800 =

Temperature offset
calibration in progress

• 0x1000 = FD

calibration in progress

• 0x2000 = D2

calibration in progress

• 0x4000 = D1

calibration in progress

• 0x8000 = Zero

calibration in progress

Table C-7

Diagnostics Object (0x66) – Instance 1 continued

Attrib
ID

Name

Data type

Service

Mem

Description

Comments

Advertising