Figure 10-10

D3 or D3 and D4 density calibration – ProLink II

Enter density of D3 fluid

Calibration in Progress

light turns green

Calibration in Progress

light turns red

D3 calibration

Close shutoff valve

downstream from sensor

Fill sensor with D3 fluid

Close

Enter density of D4 fluid

Calibration in Progress

light turns green

Calibration in Progress

light turns red

D4 calibration

Close

Done

Do Cal

Do Cal

Done

ProLink Menu >
Calibration >
Density cal – Point 3

Fill sensor with D4 fluid

ProLink Menu >
Calibration >
Density cal – Point 4