12

Alarm status 1

UINT

Get

V

A collection of status bits

• 0x0001 = NV error

(CP)

• 0x0002 = RAM error

(CP)

• 0x0004 = RTI failure
• 0x0008 = Sensor

failure

• 0x0010 =

Temperature out of
range

• 0x0020 = Calibration

failed

• 0x0040 = Other

failure

• 0x0080 = Transmitter

initializing

• 0x0100 = Not used
• 0x0200 = Not used
• 0x0400 = Simulation

mode On

• 0x0800 = Not used
• 0x1000 = Watchdog

error

• 0x2000 = Not used
• 0x4000 = Not used
• 0x8000 = Fault

13

Alarm status 2

UINT

Get

V

A collection of status bits

• 0x0001 = Not used
• 0x0002 = Not used
• 0x0004 = Not used
• 0x0008 = Not used
• 0x0010 = Density out

of range

• 0x0020 = Drive out of

range

• 0x0040 = CEM

communications error

• 0x0080 = Not used
• 0x0100 = Non-volatile

memory error (CP)

• 0x0200 = RAM error

(CP)

• 0x0400 = Sensor

failure

• 0x0800 =

Temperature out of
range

• 0x1000 = Input out of

range

• 0x2000 = Not used
• 0x4000 = Transmitter

not characterized

• 0x8000 = RTI failure

Table C-7

Diagnostics Object (0x66) – Instance 1 continued

Attrib
ID

Name

Data type

Service

Mem

Description

Comments

Advertising