Vicenet tool, see figure 10-11 – Emerson Process Management MICRO MOTION 2400S User Manual

Page 116

Advertising
background image

108

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Measurement Performance

Figure 10-11

D3 or D3 and D4 density calibration – DeviceNet tool

Enter density of

D3 fluid

Monitor status

D3 calibration

Close shutoff valve

downstream from sensor

Fill sensor with

D3 fluid

Fill sensor with

D4 fluid

D4 calibration

Done

Object: Calibration Object (0x65)
Instance: 1
Service: 0x4F

Start D3

calibration

Object: Calibration Object (0x65)
Instance: 1
Attribute ID: 15
Data type: REAL
Service: Set

Yes

Object: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 16
Bit: 7
Service: Get

Bit 7 0ff?

No

Enter density of

D4 fluid

Monitor status

Object: Calibration Object (0x65)
Instance: 1
Service: 0x50

Start D4

calibration

Object: Calibration Object (0x65)
Instance: 1
Attribute ID: 16
Data type: REAL
Service: Set

Object: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 16
Bit: 6
Service: Get

Bit 6 0ff?

Yes

No

Done