Vicenet tool, see figure 10-9 – Emerson Process Management MICRO MOTION 2400S User Manual

Page 114

Advertising
background image

106

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Measurement Performance

Figure 10-9

D1 and D2 density calibration – DeviceNet tool

Enter density of

D1 fluid

Monitor status

D1 calibration

Close shutoff valve

downstream from sensor

Fill sensor with

D1 fluid

Fill sensor with

D2 fluid

D2 calibration

Done

Object: Calibration object (0x65)
Instance: 1
Service: 0x4C

Start D1

calibration

Object: Calibration object (0x65)
Instance: 1
Attribute ID: 12
Data type: REAL
Service: Set

Yes

Object: Diagnostics object (0x66)
Instance: 1
Attribute ID: 16
Bit: 14
Service: Get

Bit 14 0ff?

No

Enter density of

D2 fluid

Monitor status

Object: Calibration object (0x65)
Instance: 1
Service: 0x4D

Start D2

calibration

Object: Calibration object (0x65)
Instance: 1
Attribute ID: 13
Data type: REAL
Service: Set

Object: Diagnostics object (0x66)
Instance: 1
Attribute ID: 16
Bit: 13
Service: Get

Bit 13 0ff?

Yes

No