DeviceNet tool – Flowmeter zero procedure

10.6

Performing density calibration

Density calibration includes the following calibration points:

•

All sensors:

-

D1 calibration (low-density)

-

D2 calibration (high-density)

•

T-Series sensors only:

-

D3 calibration (optional)

-

D4 calibration (optional)

For T-Series sensors, the optional D3 and D4 calibrations could improve the accuracy of the density
measurement. If you choose to perform the D3 and D4 calibration:

•

Do not perform the D1 or D2 calibration.

•

Perform D3 calibration if you have one calibrated fluid.

•

Perform both D3 and D4 calibrations if you have two calibrated fluids (other than air and
water).

Modify zero time if

required

Perform zero

Class: Calibration Object (0x65)
Instance: 1
Attribute ID: 3
Data type: UINT
Units: seconds
Service: Set

Class: Calibration Object (0x65)
Instance: 1
Service: 0x4B (perform or abort zero)
Value:
· 0: Abort zero calibration
· 1: Start zero calibration

Check outcome

Class: Diagnostics Object (0x66)
Instance: 1

Attribute ID 12, Bit 0x0020

Value:
· 0: Zero succeeded
· 1: Zero failed
Data type: USINT

Check zero values

Class: Calibration Object (0x65)
Instance: 1
Attribute ID 4: Zero standard deviation
Attribute ID 5: Zero offset
Data type: REAL
Units: milliseconds

Check status

Class: Diagnostics Object (0x66)
Instance: 1

Attribute ID 16, Bit 0x8000

Value:
· 0: Zero complete
· 1: Zero in progress
Data type: USINT