Emerson Process Management MICRO MOTION 2400S User Manual

Page 136

Advertising
background image

128

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Troubleshooting

Figure 11-4

Feedthrough pins

7. Using the DMM, check each pin as follows:

a.

Check between the pin and the sensor case.

b.

Check between the pin and other pins as described below:

Drive + against all other pins except Drive –

Drive – against all other pins except Drive +

Left pickoff + against all other pins except Left pickoff –

Left pickoff – against all other pins except Left pickoff +

Right pickoff + against all other pins except Right pickoff –

Right pickoff – against all other pins except Right pickoff +

RTD + against all other pins except RTD – and LLC/RTD

RTD – against all other pins except RTD + and LLC/RTD

LLC/RTD against all other pins except RTD + and RTD –

Flow tube temperature sensor

RTD + and RTD –

100

Ω

at 0 °C + 0.38675

Ω

/ °C

LLC/RTD

• T-Series sensors

RTD – and composite RTD

300

Ω

at 0 °C + 1.16025

Ω

/ °C

• CMF400 I.S. sensors

RTD – and fixed resistor

39.7–42.2

Ω

• F300 sensors

RTD – and fixed resistor

44.3–46.4

Ω

• All other sensors

RTD – and LLC

0

(1) Actual resistance values depend on the sensor model and date of manufacture. Contact Micro Motion for more detailed data.

Table 11-8

Nominal resistance ranges for flowmeter circuits continued

Circuit

Pin pairs

Nominal resistance range

(1)

Left pickoff –

Right pickoff –

Drive –

Drive +

Right pickoff +

Left pickoff +

LLC / Composite RTD /
Fixed resistor

(1)

RTD +

(1) Lead length compensator (LLC) for all sensors except T-Series, CMF400 I.S., and F300. For T-Series sensors, functions

as composite RTD. For CMF400 I.S. and F300 sensors, functions as fixed resistor.

Return for RTD, LLC,

composite RTD, or fixed resistor

Advertising