Emerson Process Management MICRO MOTION 2400S User Manual

Page 125

Advertising
background image

Configuration and Use Manual

117

Troubleshooting

Me

asu

reme

nt P

er
form

anc

e

Defaults

T

rou
b

lesho

o

ti

ng

Compensation

A009

Transmitter
Initializing/Warming
Up

Transmitter in power-up
mode

• Allow the flowmeter to warm up (approximately 30

seconds). The error should disappear once the flowmeter
is ready for normal operation.

• If alarm does not clear, make sure that the sensor is

completely full or completely empty.

• Check the sensor circuitry. See Section 11.20.

A010

Calibration Failure

Mechanical zero: The
resulting zero was greater
than 3

μ

s.

Temperature/Density Cals:
many possible causes.

• If alarm appears during a transmitter zero, ensure that

there is no flow through the sensor, then retry.

• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the

flowmeter to operation.

A011

Zero Too Low

See A10

• Ensure that there is no flow through the sensor, then retry.
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the

flowmeter to operation.

A012

Zero Too High

See A10

• Ensure that there is no flow through the sensor, then retry.
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the

flowmeter to operation.

A013

Zero Too Noisy

See A10

• Remove or reduce sources of electromechanical noise,

then retry. Sources of noise include:
- Mechanical pumps
- Pipe stress at sensor
- Electrical interference
- Vibration effects from nearby machinery

• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the

flowmeter to operation.

A014

Transmitter Failed

Many possible causes

• Cycle power to the flowmeter.
• The transmitter might need service. Contact Micro Motion.

See Section 11.3.

A016

Line RTD
Temperature
Out-Of-Range

The value computed for
the resistance of the Line
RTD is outside limits

• Check the sensor RTD circuitry. See Section 11.20.
• Verify that process temperature is within range of sensor

and transmitter.

• If the problem persists, contact Micro Motion. See

Section 11.3.

A017

Meter RTD
Temperature
Out-of-Range

The value computed for
the resistance of the
Meter/Case RTD is
outside limits

• Check the sensor RTD circuitry. See Section 11.20.
• Verify that process temperature is within range of sensor

and transmitter.

• Check the characterization. Specifically, verify the FCF

and K1 values. See Section 6.2.

• If the problem persists, contact Micro Motion. See

Section 11.3.

A020

Calibration Factors
Unentered
(FlowCal)

The flow calibration factor
and/or K1 has not been
entered since the last
master reset

• Check the characterization. Specifically, verify the FCF

and K1 values. See Section 6.2.

• If the problem persists, contact Micro Motion. See

Section 11.3.

A021

Incorrect Sensor
Type (K1)

The sensor is recognized
as a straight tube but the
K1 value indicates a
curved tube, or vice versa

• Check the characterization. Specifically, verify the FCF

and K1 values. See Section 6.2.

• Check the sensor RTD circuitry. See Section 11.20.
• If the problem persists, contact Micro Motion. See

Section 11.3.

A029

PIC/Daughterboard
Communication
Failure

Transmitter electronics
failure

• Cycle power to the flowmeter.
• Contact Micro Motion. See Section 11.3.

Table 11-2

Status alarms and remedies continued

Alarm
code

ProLink II
message

Cause

Suggested remedy