See figure 10-8 – Emerson Process Management MICRO MOTION 2400S User Manual

Page 113

Advertising
background image

Configuration and Use Manual

105

Measurement Performance

Me

asu

reme

nt P

er
form

anc

e

Defaults

T

rou
b

lesho

o

ti

ng

Compensation

Figure 10-8

D1 and D2 density calibration – ProLink II

Enter density of D1 fluid

Calibration in Progress

light turns green

Calibration in Progress

light turns red

D1 calibration

Close shutoff valve

downstream from sensor

Fill sensor with D1 fluid

Fill sensor with D2 fluid

Close

Enter density of D2 fluid

Calibration in Progress

light turns green

Calibration in Progress

light turns red

D2 calibration

Close

Done

Do Cal

Do Cal

ProLink Menu >
Calibration >
Density cal – Point 1

ProLink Menu >
Calibration >
Density cal – Point 2