Performing temperature calibration

Temperature calibration is a two-part procedure: temperature offset calibration and temperature slope
calibration. The entire procedure must be completed without interruption.

To perform temperature calibration, you must use ProLink II. See Figure 10-12.

Figure 10-12

Temperature calibration – ProLink II

Enter temperature of low-

temperature fluid

Temperature Offset calibration

Do Cal

Wait until sensor achieves

thermal equilibrium

Fill sensor with low-

temperature fluid

Calibration in Progress

light turns green

Calibration in Progress

light turns red

Close

Enter temperature of high-

temperature fluid

Temperature Slope calibration

Do Cal

Wait until sensor achieves

thermal equilibrium

Fill sensor with high-

temperature fluid

Calibration in Progress

light turns green

Calibration in Progress

light turns red

Close

Done

ProLink Menu >
Calibration >
Temp offset cal

ProLink Menu >
Calibration >
Temp slope cal