Micro Motion

®

Model 2400S Transmitters for DeviceNet

Using the Transmitter

With ProLink II

The totalizer and inventory control functions available with ProLink II are listed in Table 7-7. Note
the following:

ProLink II does not support separate resetting of the API volume totalizer and API volume
inventory. To reset these, you must reset all totalizers or all inventories.

By default, the ability to reset inventories from ProLink II is disabled. To enable it:

a.

Click

View > Preferences

.

b.

Check the

Enable Inventory Totals Reset

checkbox.

c.

Click

Apply

.

To start or stop all totalizers and inventories:

1. Click

ProLink > Totalizer Control

or

ProLink > ED Totalizer Control

(if the enhanced

density application is enabled).

2. Click the All Totals

Start

or All Totals

Stop

button.

Note: The All Totals functions are replicated in these two windows for convenience. You can start or
stop all totalizers and inventories from either window.

To reset all totalizers:

1. Click

ProLink > Totalizer Control

or

ProLink > ED Totalizer Control

(if the enhanced

density application is enabled).

2. Click the All Totals

Reset

button.

To reset all inventories:

1. Click

ProLink > Totalizer Control

or

ProLink > ED Totalizer Control

(if the enhanced

density application is enabled).

2. Click the All Totals

Reset Inventories

button.

Table 7-7

Totalizer and inventory control functions supported by ProLink II

Inventory reset

Object

Function

Disabled

Enabled

Totalizers and
inventories

Starting and stopping as a group

Totalizers

Resetting all

Resetting mass totalizer separately

Resetting volume totalizer separately

Resetting enhanced density totalizers separately

Resetting API volume totalizer separately

Not supported

Not supported

Inventories

Resetting all

Resetting mass inventory separately

Resetting volume inventory separately

Resetting enhanced density inventories separately

Resetting API volume inventory separately

Not supported

Not supported