118

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Troubleshooting

A030

Incorrect Board
Type

The loaded software is not
compatible with the
programmed board type

• Contact Micro Motion. See Section 11.3.

A031

Low Power

The transmitter is not
receiving enough power

• Check power supply to transmitter. See Section 11.4.

A032

Meter
Verification/Outputs
In Fault

Meter verification in
progress, with outputs set
to fault

• Allow the procedure to complete.
• If desired, abort the procedure and restart with outputs set

to last measured value.

A033

Sensor OK, Tubes
Stopped by
Process

No signal from LPO or
RPO, suggesting that
sensor tubes are not
vibrating

• Verify process. Check for air in the flow tubes, tubes not

filled, foreign material in tubes, or coating in tubes (see
Section 11.15).

A102

Drive Overrange/
Partially Full Tube

The drive power
(current/voltage) is at its
maximum

• Excessive drive gain. See Section 11.19.3.
• Check the sensor circuitry. See Section 11.20.
• If this is the only active alarm, it can be ignored. If desired,

reconfigure the alarm severity to Ignore (see Section 8.8).

A104

Calibration in
Progress

A calibration procedure is
in progress

• Allow the flowmeter to complete calibration.
• For zero calibration procedures, you may abort the

calibration, set the zero time parameter to a lower value,
and restart the calibration.

A105

Slug Flow

The density has exceeded
the user-defined slug
(density) limits

• See Section 11.14.

A107

Power Reset
Occurred

The transmitter has been
restarted

• No action required.
• If desired, reconfigure the alarm severity to Ignore (see

Section 8.8).

A116

API: Temperature
Outside Standard
Range

The process temperature
is outside API-defined
extrapolation limits

• Verify process.
• Verify API reference table and temperature configuration.

See Section 8.13.

A117

API: Density
Outside Standard
Range

The process density is
outside API-defined
extrapolation limits

• Verify process.
• Verify API reference table and density configuration. See

Section 8.13.

A120

ED: Unable to Fit
Curve Data

The configured values for
density curves do not
meet accuracy
requirements

• Verify enhanced density configuration. See Section 8.14.

A121

ED: Extrapolation
Alarm

Enhanced density
calculations are outside
the configured data range

• Verify process temperature.
• Verify process density.
• Verify enhanced density configuration. See Section 8.14.

A131

Meter
Verification/Outputs
at Last Value

Meter verification in
progress, with outputs set
to last measured value

• Allow the procedure to complete.
• If desired, abort the procedure and restart with outputs set

to fault.

A132

Simulation Mode
Active

Simulation mode is
enabled

• Disable simulation mode. See Section 11.10.

A133

PIC UI EEPROM
Error

EEPROM data on the user
interface module is corrupt

• Contact Micro Motion. See Section 11.3.

Table 11-2

Status alarms and remedies continued

Alarm
code

ProLink II
message

Cause

Suggested remedy