Pressure Compensation and Temperature Compensation

9.2.2

Pressure correction factors

When configuring pressure compensation, you must provide the flow calibration pressure – the
pressure at which the flowmeter was calibrated (which therefore defines the pressure at which there
will be no effect on the calibration factor). Enter

20 PSI

unless the calibration document for your

sensor indicates a different calibration pressure.

Two additional pressure correction factors may be configured: one for flow and one for density. These
are defined as follows:

Flow factor – the percent change in the flow rate per psi

Density factor – the change in fluid density, in g/cm

3

/psi

Not all sensors or applications require pressure correction factors. For the pressure correction values
to be used, obtain the pressure effect values from the product data sheet for your sensor, then reverse
the signs (e.g., if the pressure effect is

0.000004

, enter a pressure correction factor of

–0.000004

).

9.2.3

Configuration

To enable and configure pressure compensation:

With ProLink II, see Figure 9-1.

With a DeviceNet tool, see Figure 9-2.

Figure 9-1

Configuring pressure compensation with ProLink II

Enter Flow factor

Configure

Enter Density factor

Enter Cal pressure

Set up output

assembly

(2)

Enter External

Pressure

Enable External Pressure

Compensation

Enable

Apply

Enter Pressure units

Set measurement unit

(1)

Yes

Done

Apply

View >
Preferences

ProLink >
Configuration >
Pressure

ProLink >
Configuration >
Pressure

Apply

Apply

Use static

pressure value?

No

(1) Pressure measurement unit must be configured to match pressure

unit used by external device. See Section 6.3.

(2) See Section 9.4.