Llustrated in figure 10-2. f – Emerson Process Management MICRO MOTION 2400S User Manual

Page 101

Advertising
background image

Configuration and Use Manual

93

Measurement Performance

Me

asu

reme

nt P

er
form

anc

e

Defaults

T

rou
b

lesho

o

ti

ng

Compensation

Figure 10-2

Meter verification procedure – Display menu

Select

OFF-LINE MAINT

Scroll and Select simultaneously

for 4 seconds

Scroll

Select

Scroll

SENSOR VERFY

OUTPUTS

STOP MSMT/YES?

. . . . . . . . . . . . . . . x%

(2)

UNSTABLE FLOW

(1)

Select

PASS

ABORT

CAUTION

Scroll

Scroll

RERUN/YES?

Yes

Scroll

No

Scroll

Select

Correct condition

Correct condition

Select

Scroll

Scroll

SENSOR EXIT

Scroll

Choose output setting

(1) Either Unstable Flow or Unstable Drive Gain may

be displayed, indicating that the standard deviation
of the flow or drive gain is outside limits.

(2) Represents the percentage completion of the

procedure.