12 status alarms, Status alarms, Table 11-2. be – Emerson Process Management MICRO MOTION 2400S User Manual

Page 124: Table 11-2.), Section 11.12, The alarm (see section 11.12)

Advertising
background image

116

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Troubleshooting

11.12

Status alarms

Status alarm codes are reported on the LCD panel (for transmitters that have a display), and status
alarms can be viewed with ProLink II or a DeviceNet tool (see Section 7.6). All possible status alarms
are listed in Table 11-2, along with the ProLink II message, possible causes, and suggested remedies.

You may find it useful to acknowledge all alarms before beginning the troubleshooting procedures.
This will remove inactive alarms from the list and allow you to focus on active alarms.

Table 11-2

Status alarms and remedies

Alarm
code

ProLink II
message

Cause

Suggested remedy

A001

(E)EPROM
Checksum Error
(CP)

An uncorrectable
checksum mismatch has
been detected

• Cycle power to the flowmeter.
• The flowmeter might need service. Contact Micro Motion.

See Section 11.3.

A002

RAM Error (CP)

ROM checksum error or a
RAM location cannot be
written to

• Cycle power to the flowmeter.
• The flowmeter might need service. Contact Micro Motion.

See Section 11.3.

A003

Sensor Failure

Continuity failure of drive
circuit, LPO, or RPO, or
LPO-RPO mismatch when
driving

• Check for slug flow. See Section 11.14.
• Check the test points. See Section 11.19.
• Check the sensor circuitry. See Section 11.20.
• Check sensor tubes for plugging.
• If the problem persists, contact Micro Motion. See

Section 11.3.

A004

Temperature
Sensor Failure

Combination of A016 and
A017

• Check the sensor RTD circuitry. See Section 11.20.
• Verify that process temperature is within range of sensor

and transmitter.

• If the problem persists, contact Micro Motion. See

Section 11.3.

A005

Input Overrange

The measured flow has
exceeded the maximum
flow rate of the sensor
(

Δ

T > 200

μ

s)

• If other alarms are present (typically, A003, A006, A008,

A102, or A105), resolve those alarm conditions first. If the
A005 alarm persists, continue with the suggestions here.

• Verify process and check for slug flow. See Section 11.14.
• Check the test points. See Section 11.19.
• Check the sensor circuitry. See Section 11.20.
• Check the sensor tubes for erosion. See Section 11.15.
• If the problem persists, contact Micro Motion. See

Section 11.3.

A006

Not Configured

Combination of A020 and
A021

• Check the characterization. Specifically, verify the FCF

and K1 values. See Section 6.2.

• If the problem persists, contact Micro Motion. See

Section 11.3.

A008

Density Overrange

The measured density has
exceeded 0–10 g/cm

3

• If other alarms are present (typically, A003, A006, A102, or

A105), resolve those alarm conditions first. If the A008
alarm persists, continue with the suggestions here.

• Verify process. Check for air in the flow tubes, tubes not

filled, foreign material in tubes, or coating in tubes (see
Section 11.15).

• Check for slug flow. See Section 11.14.
• Check the sensor circuitry. See Section 11.20.
• Verify calibration factors in transmitter configuration. See

Section 6.2.

• Check the test points. See Section 11.19.
• If the problem persists, contact Micro Motion. See

Section 11.3.