Model 2400S Transmitters for DeviceNet

™

Optional Configuration

API reference tables

Reference tables are organized by reference temperature, CTL derivation method, liquid type, and
density unit. The table selected here controls all the remaining options.

•

Reference temperature:

-

If you specify a 5x, 6x, 23x, or 24x table, the default reference temperature is 60

°

F, and

cannot be changed.

-

If you specify a 53x or 54x table, the default reference temperature is 15

°

C. However, you

can change the reference temperature, as recommended in some locations (for example, to
14.0 or 14.5

°

C).

•

CTL derivation method:

-

If you specify an odd-numbered table (5, 23, or 53), CTL will be derived using method 1
described above.

-

If you specify an even-numbered table (6, 24, or 54), CTL will be derived using method 2
described above.

•

The letters A, B, C, or D that are used to terminate table names define the type of liquid that the
table is designed for:

-

A tables are used with generalized crude and JP4 applications.

-

B tables are used with generalized products.

-

C tables are used with liquids with a constant base density or known thermal expansion
coefficient.

-

D tables are used with lubricating oils.

•

Different tables use different density units:

-

Degrees API

-

Relative density (SG)

-

Base density (kg/m

3

)