Emerson Process Management MICRO MOTION 2400S User Manual

Page 164

Advertising
background image

156

Micro Motion

®

Model 2400S Transmitters for DeviceNet

Device Profile

55

Meter
verification abort
code

USINT

Get

V

The reason the meter
verification routine
aborted

• 0 = No error
• 1 = Manual abort
• 2 = Watchdog timeout
• 3 = Frequency drift
• 4 = High peak drive

voltage

• 5 = High drive current

standard deviation

• 6 = High drive current

mean

• 7 = Drive loop

reported error

• 8 = High Delta T

standard deviation

• 9 = High Delta T

value

• 10 = State running
• 11 = Verification

complete

• 12 = Wrong

verification enable

• 13 = No factory air

verification

• 14 = No factory water

verification

• 15 = Parameters not

set

56

Meter
verification
algorithm state
at abort

USINT

Get

V

The state of the meter
verification routine when it
aborted

1–18

57

Meter
verification
percent
complete

USINT

Get

V

The progress of the meter
verification routine

%

58

Meter
verification
outputs state

USINT

Set

NV

The state of the outputs
when the meter
verification routine is
running

• 0 = Last value
• 1 = Fault

59

Meter
verification
stiffness limit

REAL

Set

NV

The setpoint of the
stiffness limit. Represents
percentage.

Unitless

60

Meter
verification
validation
counter

UINT

Get

NV

Indicates the number of
times the meter
verification routine has
successfully completed

61

Meter
verification inlet
stiffness out of
limits

USINT

Get

V

Is the inlet stiffness out of
limits?

• 0 = No
• 1 = Yes

62

Meter
verification
outlet stiffness
out of limits

USINT

Get

V

Is the outlet stiffness out
of limits?

• 0 = No
• 1 = Yes

Table C-7

Diagnostics Object (0x66) – Instance 1 continued

Attrib
ID

Name

Data type

Service

Mem

Description

Comments