5.4.4. RANGE UNITS

The GFC7001T/GFC7001TM can display concentrations in parts per million (10

6

mols

per mol, PPM) or milligrams per cubic meter (mg/m

3

, MG). Changing units affects all

of the display, COMM port and DAS values for all reporting ranges regardless of the
analyzer’s range mode. To change the concentration units:

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

In order to avoid a reference temperature bias, the analyzer must be
recalibrated after every change in reporting units.

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

Concentrations displayed in mg/m

3

and ug/m

3

use 0

° C

and 760 mmHg for

Standard Temperature and Pressure (STP). Consult your local regulations
for the STP used by your agency.
(Example: US EPA uses 25

o

C as the reference temperature).

Once the Units of Measurement have been changed from volumetric (ppb or
ppm) to mass units (µg/m

3

or mg/m

3

) the analyzer MUST be recalibrated, as

the “expected span values” previously in effect will no longer be valid.
Simply entering new expected span values without running the entire
calibration routine IS NOT sufficient.
This will also counteract any discrepancies between STP definitions.