Calibration Procedures Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

218

9.7. CALIBRATION OF OPTIONAL SENSORS

This section provides the calibration setup and procedures for the O

2

Sensor and the CO

2

Sensor options. Refer to any addenda accompanying this manual for custom
configuration information.

9.7.1. O

2

SENSOR CALIBRATION

Presented here are first the setup and then the calibration steps for the O

2

Sensor.

9.7.1.1. O2 PNEUMATICS CONNECTIONS

The pneumatic connections for calibrating are as follows:

Figure 9-7:

O

2

Sensor Calibration Set Up

O

2

SENSOR ZERO GAS: TAI recommends using pure N

2

when calibration the zero

point of your O

2

sensor option.

O

2

SENSOR SPAN GAS: TAI recommends using 20.8% O

2

in N

2

when calibration the

span point of your O

2

sensor option (See Table 3-12).

9.7.1.2. SET O2 SPAN GAS CONCENTRATION

Set the expected O

2

span gas concentration.

This should be equal to the percent concentration of the O

2

span gas of the selected

reporting range (default factory setting = 20.8%; the approximate O

2

content of ambient

air).