Precision calibration, Auditing procedure, Precision calibration procedures – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 232: Calibration audit

Advertising
background image

EPA Calibration Protocol Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

232

10.3. PRECISION CALIBRATION

Calibration must be performed with a calibrator that meets all conditions specified in
QA Handbook

1

Vol II Part 1, App 15, Sec. 4.4 & 5.4. The user should be sure that all

flow meters are calibrated under the conditions of use against a reliable standard. All
volumetric flow rates should be corrected to 25

o

C (77

o

F) and 760 mm-Hg (29.92in–Hg).

Make sure the calibration system can supply the range of the concentration at a
sufficient flow over the whole range of concentration that will be encountered during
calibration.
All operational adjustments to the GFC7001T should be completed prior to the
calibration. The following software features must be set into the desired state before
calibration.
 If the instrument will be used for more than one range, it should be calibrated

separately on each applicable range.

 Automatic temperature/pressure compensation should be enabled. See Section 5.7.
 Alternate units, make sure ppm units are selected for EPA monitoring. See Section

5.4.4.

The analyzer should be calibrated on the same range used for monitoring.

10.3.1. PRECISION CALIBRATION PROCEDURES

To perform a precision calibration during the instrument set up, the input sources of zero
air and sample gas and procedures should conform to those described in Section 9.2 for
analyzers with no valve options or with an IZS valve option installed and Section 9.3 for
analyzers with Z/S options installed.

10.4. AUDITING PROCEDURE

An audit is an independent assessment of the accuracy of data. Independence is
achieved by having the audit made by an operator other than the one conducting the
routine field measurements and by using audit standards and equipment different from
those routinely used in monitoring. The audit should be a true assessment of the
measurement process under normal operations without any special preparation or
adjustment of the system. Routine quality control checks conducted by the operator are
necessary for obtaining and reporting good quality data, but they are not considered part
of the auditing procedure. Audits are recommended once per quarter, but frequency
may be determined by applicable regulations and end use of the data.
Refer to The Q.A. Handbook

1

Volume II, Part 1 Section 16 (for a more detailed

description).

10.4.1. CALIBRATION AUDIT

A calibration audit consists of challenging the GFC7001T/GFC7001TM with known
concentrations of CO. The difference between the known concentration and the
analyzer response is obtained, and an estimate of the analyzer's accuracy is determined.
The recommended audit schedule depends on the purpose for which the monitoring data
are being collected. For example, Appendix A, 40 CFR 58

5

requires that each analyzer

in State and Local Air Monitoring Network Plan (SLAMS) be audited at least once a

Advertising