References – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 235

Advertising
background image

EPA Calibration Protocol Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

235

 Failed IR detector, GFC Wheel or Sync/Demod Board
 Contaminated optical bench or sample lines

10.6. REFERENCES

1

Quality Assurance Handbook for Air Pollution Measurement Systems Volume II: Part
1 - Ambient Air Quality Monitoring Program Quality System Development - EPA-
454/R-98-004 - August 1998. United States Environmental Protection Agency -
Office of Air Quality Planning and Standards

2

CFR Title 40: Protection of Environment - PART 53—AMBIENT AIR MONITORING
REFERENCE AND EQUIVALENT METHODS:
- 53.20 General provisions.
- 53.23 Test procedures.

3

CFR Title 40: Protection of Environment - PART 50—NATIONAL PRIMARY AND
SECONDARY AMBIENT AIR QUALITY STANDARDS: Appendix C to Part 50—
Measurement Principle and Calibration Procedure for the Measurement of Carbon
Monoxide in the Atmosphere (Non-Dispersive Infrared Photometry)

4

Quality Assurance Handbook for Air Pollution Measurement Systems - Volume II,

Ambient Air Specific Methods, EPA-600/4-77-027a, 1977.

5

CFR Title 40: Protection of Environment - AMBIENT AIR QUALITY SURVEILLANCE

Advertising