CommunicationsTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

161

To assign or reset the status flag bit assignments, press:

SAMPLE

RANGE=50.0 PPM CO= XX.XX

<TST

TST> CAL

SETUP

DEL deletes the

character currently

inside the cursor

brackets.

Press the [?] button repeatedly to cycle

through the available character set: 0-9

NOTE: Values of A-F can also be set

but are meaningless.

The <CH and CH>

buttons move the

cursor brackets “[ ]”

left and right along the

bit string.

SETUP X.X

HESSEN STATUS FLAGS

<SET

EDIT

EXIT

SETUP X.X

COMMUNICATIONS MENU

ID

HESN

COM1 COM2

EXIT

SETUP X.X

HESSEN VARIATION:TYPE1

<SET

SET>

EDIT

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG

EXIT

Continue pressing NEXT until ...

SETUP X.X

SOURCE WARNING:0004

NEXT

EDIT PRNT EXIT

SETUP X.X

BOX TEMP WARNING:[0]008

<CH CH>

DEL

[0]

ENTR EXIT

SETUP X.X

SAMPLE TEMP WARNING:0020

PREV NEXT

EDIT

PRNT EXIT

Continue pressing NEXT until desired

flag message is displayed

EXIT discards the

new setting.

ENTR accepts the

new setting.

6.7.3.5. INSTRUMENT ID

Each instrument on a Hessen Protocol network must have a unique identifier (ID
number). If more than one GFC7001T/GFC7001TM analyzer is on the Hessen network,
refer to Section 5.7.1 for information and to customize the ID of each.