Getting StartedTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

66

Table 3-9: Zero/Span Valve Operating States for Option 50B

MODE VALVE

CONDITION

SAMPLE

(Normal

State)

Sample/Cal

Open to SAMPLE inlet

Zero/Span

Open to IZS inlet

Shutoff Valve

Closed

ZERO CAL

Sample/Cal

Open to ZERO/SPAN valve

Zero/Span

Open to IZS inlet

Shutoff Valve

Closed

SPAN CAL

Sample/Cal

Open to ZERO/SPAN valve

Zero/Span

Open to SHUTOFF valve

Shutoff Valve

Open to PRESSURE SPAN Inlet

3.3.2.6. PNEUMATIC CONNECTIONS FOR ZERO SCRUBBER/PRESSURIZED

SPAN OPTION

Figure 3-24:

Pneumatic Connections – Zero Scrubber/Pressurized Span Calibration Valves (Opt 50E)

S

AMPLE

G

AS

S

OURCE

Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not
be more than 2 meters long.
 Maximum pressure of any gas at the sample inlet should not exceed 1.5 in-hg above

ambient pressure and ideally should equal ambient atmospheric pressure.

 In applications where the sample gas is received from a pressurized manifold, a vent

must be placed on the sample gas before it enters the analyzer.

C

ALIBRATION

G

AS

S

OURCES

S

PAN

G

AS

:

 Attach a gas line from the pressurized source of calibration gas (e.g. a bottle of

NIST-SRM gas) to the span inlet.