SetupTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

113

referred to in many places of the maintenance and trouble-shooting sections of this
manual.
The various operating modes available under the DIAG menu are:

Table 5-4: Diagnostic Mode (DIAG) Functions

DIAG SUBMENU

SUBMENU FUNCTION

Front Panel Mode

Indicator

MANUAL

SECTION

SIGNAL I/O

Allows observation of all digital and analog signals in
the instrument. Allows certain digital signals such as
valves and heaters to be toggled ON and OFF.

DIAG SIGNAL

I/O

5.9.1,

12.1.3 and

12.5.8.1

ANALOG OUTPUT

When entered, the analyzer performs an analog
output step test. This can be used to calibrate a
chart recorder or to test the analog output accuracy.

DIAG ANALOG

OUTPUT

5.9.2 and

12.5.8.2

ANALOG I/O

CONFIGURATION

This submenu allows the user to configure the
analyzer’s analog output channels, including
choosing what parameter will be output on each
channel. Instructions that appear here allow
adjustment and calibration of the voltage signals
associated with each output as well as calibration of
the analog to digital converter circuitry on the
motherboard.

DIAG ANALOG

I/O

CONFIGURATI

ON

5.9.3

ELECTRICAL

TEST

When activated, the analyzer performs an electrical
test, which generates a voltage intended to simulate
the measure and reference outputs of the
SYNC/DEMOD board to verify the signal handling
and conditioning of these signals.

DIAG

ELECTRICAL

TEST

5.9.4 and

12.5.7.2

DARK

CALIBRATION

1

Disconnects the preamp from synchronous
demodulation circuitry on the SYNC/DEMOD PCA to
establish the dark offset values for the measure and
reference channel.

DIAG DARK

CALIBRATION

5.9.5 and

9.6.1

PRESSURE

CALIBRATION

1

Allows the user to calibrate the sample pressure
sensor.

DIAG

PRESSURE

CALIBRATION

5.9.6 and

9.6.2

FLOW

CALIBRATION

1

This function is used to calibrate the gas flow output
signals of sample gas and ozone supply.

DIAG FLOW

CALIBRATION

5.9.7 and

9.6.3

TEST CHAN

OUTPUT

Selects one of the available test channel signals to
output over the A4 analog output channel.

DIAG TEST

CHAN OUTPUT

5.9.8 and

12.5.8.2

1

These settings are retained after exiting DIAG mode.