Removing/replacing the gfc wheel – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 282

Advertising
background image

Troubleshooting and ServiceTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

282

12.6.2. REMOVING/REPLACING THE GFC WHEEL

When removing or replacing the GFC Wheel it is important to perform the disassembly
in the following order to avoid damaging the components:
1. Turn off the analyzer.
2. Remove the top cover.
3. Open the instrument’s hinged front panel.
4. Locate the GFC Wheel/motor assembly. See Figure 3-6.
5. Unplug the following electronic components:

 The GFC Wheel housing temperature sensor
 GFC Wheel heater
 GFC Wheel motor power supply

SAFETY SHIELD

SYNCHRONOUS MOTOR

SOURCE ASSEMBLY

THERMISTOR

HEATER

Figure 12-15: Opening the GFC Wheel Housing

6. Remove the two (2) screws holding the opto-pickup printed circuit assembly to the

GFC Wheel housing.