258

12.1.4.2. SYNC DEMODULATOR STATUS LEDS

Two LEDs located on the Sync/Demod Board and are there to make it obvious that the
GFC Wheel is spinning and the synchronization signals are present:

Table 12-3: Sync/Demod Board Status Failure Indications

LED

FUNCTION

FAULT STATUS

INDICATED FAILURE(S)

D1

M/R Sensor Status

(Flashes slowly)

LED is stuck

ON or OFF

GFC Wheel is not turning
M/R Sensor on Opto-Pickup Board failed
Sync/Demod Board failed
JP 4 Connector/Wiring faulty
Failed/Faulty +5 VDC Power Supply (PS1)

D2

Segment Sensor

Status

(Flashes quickly)

LED is stuck

ON or OFF

GFC Wheel is not turning
Segment Sensor on Opto-Pickup Board failed
Sync/Demod Board failed
JP 4 Connector/Wiring faulty
Failed/Faulty +5 VDC Power Supply (PS1)

D1 – M/R Sensor Status

D2 – Segment Sensor Status

JP4 Connector to Opto-Pickup

Board

Figure 12-4:

Sync/Demod Board Status LED Locations