Getting StartedTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

62

 Maximum pressure of any gas at the sample inlet should not exceed 1.5 in-hg above

ambient pressure and ideally should equal ambient atmospheric pressure.

 In applications where the sample gas is received from a pressurized manifold, a vent

must be placed on the sample gas before it enters the analyzer.

C

ALIBRATION

G

AS

S

OURCES

The source of calibration gas is also attached to the SAMPLE inlet, but only when a
calibration operation is actually being performed. The calibration gases should be
available to the analyzer at the same pressure as the sample gas and with the same flow
rate (800 cm

3

/min).

E

XHAUST

O

UTLET

Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line
should be:
 PTEF tubing; minimum O.D 1/4”;
 A maximum of 10 meters long;
 Vented outside the GFC7001T/GFC7001TM Analyzer’s enclosure.

3.3.2.2. PNEUMATIC LAYOUT FOR BASIC CONFIGURATION

Figure 3-20:

GFC7001T/GFC7001TM Internal Gas Flow