Teledyne analytical instruments 49 – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 49

Advertising
background image

Getting StartedTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

49

Table 3-6: Status Output Signals

REAR PANEL

LABEL

STATUS

DEFINITION

CONDITION

1

SYSTEM OK

ON if no faults are present.

2 CONC

VALID

OFF any time the HOLD OFF feature is active, such as during calibration or when
other faults exist possibly invalidating the current concentration measurement
(example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid.

3 HIGH

RANGE

ON if unit is in high range of either the DUAL or AUTO range modes.

4 ZERO

CAL

ON whenever the instrument’s ZERO point is being calibrated.

5 SPAN

CAL

ON whenever the instrument’s SPAN point is being calibrated.

6 DIAG

MODE

ON whenever the instrument is in DIAGNOSTIC mode.

7 CO

2

CAL

If this analyzer is equipped with an optional CO

2

sensor, this Output is ON when that

sensor is in calibration mode.
Otherwise this output is unused.

8 O

2

CAL

If this analyzer is equipped with an optional O

2

sensor, this Output is ON when that

sensor is in calibration mode.
Otherwise this output is unused.

D

EMITTER BUS

The emitters of the transistors on pins 1-8 are bussed together.

SPARE

+

DC POWER

+ 5 VDC, 300 mA source maximum.

Digital Ground

The ground level from the analyzer’s internal DC power supplies.

3.3.1.6. CONNECTING THE CONTROL INPUTS

To remotely activate the zero and span calibration modes, several digital control inputs
are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear
panel.
There are two methods for energizing the control inputs. The internal +5V available
from the pin labeled “+” is the most convenient method. However, if full isolation is
required, an external 5 VDC power supply should be used.

Advertising