Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 129

Advertising
background image

SetupTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

129

An alternative method for measuring the output of the Current Loop converter is to
connect a 250 ohm

1% resistor across the current loop output in lieu of the current

meter (see Figure 3-9 for pin assignments and diagram of the analog output connector).
This allows the use of a voltmeter connected across the resistor to measure converter
output as VDC or mVDC.

V

+DC Gnd

Figure 5-6:

Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels

In this case, follow the procedure above but adjust the output for the following values:

Table 5-8:

Current Loop Output Check

% FS

Voltage across Resistor for

2-20 mA

Voltage across

Resistor for 4-20 mA

0

500 mVDC

1000 mVDC

100

5000 mVDC

5000 mVDC

Advertising