Flow rate control – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 295

Advertising
background image

Theory of Operation Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

295

The sample gas is to be made available to the analyzer at a pressure slightly above
ambient but sufficient to allow for a proper flow though the sample lines. The internal
sample system has been designed with minimal restriction allowing sufficient flow at
very low pressure. The input pressure must be reduced and regulated from the process
pressure using a pressure regulator, needle valve, restrictor, critical orifice or any
combination thereof before entering the analyzer. .

Figure 13-7:

Internal Pneumatic Flow – Basic Configuration

13.3. FLOW RATE CONTROL

To maintain a constant flow rate of the sample gas through the instrument, the
GFC7001T/GFC7001TM uses a special flow control assembly located in the exhaust
gas line. In instruments with the O

2

sensor installed, a second flow control assembly is

located between the O

2

sensor assembly and the exhaust port. These assemblies consist

of:
 A critical flow orifice.
 Two o-rings: Located just before and after the critical flow orifice, the o-rings seal

the gap between the walls of assembly housing and the critical flow orifice.

 A spring: Applies mechanical force needed to form the seal between the o-rings, the

critical flow orifice and the assembly housing.

13.3.1.1. CRITICAL FLOW ORIFICE

The most important component of this flow control assembly is the critical flow orifice.

Advertising