Theory of Operation Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

306

13.4.4.3. ZERO/SPAN VALVE OPTIONS

Any zero/span/shutoff valve options installed in the analyzer are controlled by a set of
electronic switches located on the relay board. These switches, under CPU control,
supply the +12VDC needed to activate each valve’s solenoid.

13.4.4.4. IR SOURCE

The relay board supplies a constant 11.5VDC to the IR Source. Under normal operation
the IR source is always on.

13.4.4.5. STATUS LEDS

Eight LEDs are located on the analyzer’s relay board to show the current status on the
various control functions performed by the relay board. They are listed on Table 13-4.

Table 13-4: Relay Board Status LEDs

LED

COLOR

FUNCTION

STATUS WHEN LIT

STATUS WHEN UNLIT

D1

RED

Watch Dog Circuit

Cycles On/Off every 3 seconds under direct control of the analyzer’s
CPU.

D2

YELLOW

Wheel Heater

HEATING

NOT HEATING

D3

YELLOW

Bench Heater

HEATING

NOT HEATING

D4 YELLOW

Spare

N/A

N/A

D5 GREEN

Sample/Cal Gas Valve

Option

Valve Open to CAL GAS FLOW

Valve Open to SAMPLE Gas Flow

D6

GREEN

Zero/Span Gas Valve Option

Valve Open to SPAN GAS FLOW

Valve Open to ZERO GAS FLOW

D7 GREEN Shutoff

Valve

Option

Valve Open to CAL GAS FLOW

Valve CLOSED to CAL GAS

FLOW

D8

GREEN

IR SOURCE

Source ON

Source OFF