Gas filter correlation, Teledyne analytical instruments 291 – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 291

Advertising
background image

Theory of Operation Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

291

IR

Source

IR Beam

Sample Chamber

Band-Pass Filter

Photo-Detector

Figure 13-1:

Measurement Fundamentals

Upon exiting the sample cell, the beam shines through a band-pass filter that allows only
light at a wavelength of 4.7 µm to pass. Finally, the beam strikes a solid-state photo-
detector that converts the light signal into a modulated voltage signal representing the
attenuated intensity of the beam.

13.2.1. GAS FILTER CORRELATION

Unfortunately, water vapor absorbs light at 4.7 µm too. To overcome the interfering
effects of water vapor the GFC7001T/GFC7001TM adds another component to the IR
light path called a Gas Filter Correlation (GFC) Wheel.

Measurement Cell

(Pure N

2

)

Reference Cell

(N

2

with CO)

Figure 13-2:

GFC Wheel

13.2.1.1. THE GFC WHEEL

A GFC Wheel is a metallic wheel into which two chambers are carved. The chambers
are sealed on both sides with material transparent to 4.7 µm IR radiation creating two
airtight cavities. Each cavity is mainly filled with composed gases. One cell is filled
with pure N

2

(the measurement cell)

.

The other is filled with a combination of N

2

and a

high concentration of CO (the reference cell).