To access and navigate the VARS menu, use the following button sequence.

SETUP X.X 4) CLOCK_ADJUST=0 Sec/Day

PREV

NEXT

JUMP

EDIT ENTR EXIT

SETUP X.X CLOCK_ADJUST=0 Sec/Day

+

0

0

ENTR EXIT

Enter sign and number of

seconds per day the clock

gains (-) or loses(+).

SETUP X.X

DYN_SPAN=OFF

OFF

ENTR EXIT

SETUP X.X

DYN_ZERO=OFF

OFF

ENTR EXIT

Toggle this button to turn

the Dynamic Zero

calibration feature ON/

OFF.

Concentration display

continuously cycles

through all gasses.

SAMPLE

RANGE=500.0 PPM CO= XXXX

<TST

TST> CAL

SETUP

SETUP X.X

0) DAS_HOLD_OFF=15.0 Minutes

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS

DIAG

EXIT

SETUP X.X

ENTER PASSWORD:818

8

1

8

ENTR

EXIT

Toggle these

buttons to enter

the correct

PASSWORD.

In all cases:

EXIT discards the new

setting.

ENTR accepts the

new setting.

SETUP X.X

2) DYN_ZERO=OFF

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

3) DYN_SPAN=OFF

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

DAS_HOLD_OFF=15.0 Minutes

1

5

.0

ENTR EXIT

Toggle these buttons to set

the iDAS HOLDOFF time

period in minutes

(MAX = 20 minutes).

Toggle this button to turn

the Dynamic Span

calibration feature ON/

OFF.

SETUP X.X

1) CONC_PRECISION=AUTO

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

CONC_PRECISION=AUTO

AUTO

1

2

3

4

ENTR EXIT

SETUP X.X

5) STABIL_GAS=CO

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

STABIL_GAS=O2

CO

CO2

O2

ENTR EXIT

Use these buttons to select

which gas will be reported

by the sTABIL test

function.

(O

2

is only available if the

optional O

2

sensor is

installed)

Use these buttons to select

the precision of the o33

concentration display.

Press NEXT for additional

VARS; press NEXT or

PREV to move back and

forth throughout the list of

VARS.

5.9. SETUP DIAG: DIAGNOSTICS FUNCTIONS

A series of diagnostic tools is grouped together under the SETUPMOREDIAG
menu, as these parameters are dependent on firmware revision (see Appendix A). These
tools can be used in a variety of troubleshooting and diagnostic procedures and are