Teledyne analytical instruments 87 – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 87

Advertising
background image

OverviewTeledyne API – Model T300/T300M CO AnalyzerTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

87

Figure 4-2:

Viewing GFC7001T/GFC7001TM Test Functions


I

MPORTANT

I

MPACT ON

R

EADING OR

D

ATA

A value of “XXXX” displayed for any of the TEST functions indicates an
out-of-range reading or the analyzer’s inability to calculate it. All
pressure measurements are represented in terms of absolute pressure.
Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It
decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors
such as air conditioning and passing storms can cause changes in the
absolute atmospheric pressure.

Table 4-2:

Test Functions Defined

PARAMETER DISPLAY

TITLE UNITS

MEANING

Stability

STABIL

PPB

3

, PPM

UGM

3

, MGM

Standard deviation of CO concentration readings. Data points are
recorded every ten seconds using the last 25 data points. This
function can be reset to show O

2

or CO

2

stability in instruments with

those sensor options installed.

Range

RANGE

RANGE1

1

RANGE2

1

PPB, PPM,

UGM, MGM

The full scale limit at which the reporting range of the analyzer is
currently set. THIS IS NOT the Physical Range of the instrument.
See Section 5.4.1 for more information.