Troubleshooting and ServiceTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

254

values are still within the acceptable range but have significantly changed from the
measurement recorded on the factory data sheet may also indicate a failure.

Note

A worksheet has been provided in Appendix C to assist in recording the
value of these test functions. This worksheet also includes expected
values for the various test functions.

The following table contains some of the more common causes for these values to be out
of range.

Table 12-2: Test Functions - Indicated Failures

TEST

FUNCTIONS

(

As Displayed)

INDICATED FAILURE(S)

TIME

Time of day clock is too fast or slow.
To adjust, see Section 5.6.
Battery in clock chip on CPU board may be dead.

RANGE

Incorrectly configured measurement range(s) could cause response problems with a Data logger or chart
recorder attached to one of the analog output.
If the Range selected is too small, the recording device will over range.
If the Range is too big, the device will show minimal or no apparent change in readings.

STABIL

Indicates noise level of instrument or CO concentration of sample gas (see Section 12.4.2 for causes).

CO MEAS

&

CO REF

If the value displayed is too high the IR Source has become brighter. Adjust the variable gain potentiometer on
the sync/demod board (see Section 12.5.7.1).
If the value displayed is too low or constantly changing and the CO REF is OK:

Failed multiplexer on the motherboard

Failed sync/demod board

Loose connector or wiring on sync/demod board

If the value displayed is too low or constantly changing and the CO REF is bad:

GFC Wheel stopped or rotation is too slow

Failed sync/demod board IR source

Failed IR source

Failed relay board

 Failed

I

2

C bus

Failed IR photo-detector

MR Ratio

When the analyzer is sampling zero air and the ratio is too low:

The reference cell of the GFC Wheel is contaminated or leaking.

The alignment between the GFC Wheel and the segment sensor, the M/R sensor or both is incorrect.

Failed sync/demod board

When the analyzer is sampling zero air and the ratio is too high:

Zero air is contaminated

Failed IR photo-detector

PRES

See Table 12-1 for SAMPLE PRES WARN.

SAMPLE FL

Check for gas flow problems (see Section 12.2).

SAMP TEMP

SAMPLE TEMP should be close to BENCH TEMP. Temperatures outside of the specified range or oscillating
temperatures are cause for concern.

BENCH

TEMP

Bench temp control improves instrument noise, stability and drift. Temperatures outside of the specified range
or oscillating temperatures are cause for concern. Table 12-1 for BENCH TEMP WARNING.

WHEEL

TEMP

Wheel temp control improves instrument noise, stability and drift. Outside of set point or oscillating
temperatures are causes for concern. See Table 12-1 for WHEEL TEMP WARNING.