Troubleshooting and ServiceTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

283

7. Carefully remove the opto-pickup printed circuit assembly.

Opto-Pickup

Figure 12-16: Removing the Opto-Pickup Assembly

8. Remove the three (3) screws holding the GFC Wheel motor/heat sink assembly to

the GFC Wheel housing.

9. Carefully remove the GFC Wheel motor/heat sink assembly from the GFC Wheel

housing.

GFC WHEEL HOUSING

Figure 12-17: Removing the GFC Wheel Housing