140

number on the front panel display. For example, if quiet mode (01), computer mode
(02) and Multidrop-Enabled mode (32) are selected, the analyzer would display a
combined MODE ID of 35.

Table 6-1: COMM Port Communication Modes

MODE

1

ID

DESCRIPTION

QUIET

1

Quiet mode suppresses any feedback from the analyzer (such as warning messages) to
the remote device and is typically used when the port is communicating with a computer
program where such intermittent messages might cause communication problems.
Such feedback is still available but a command must be issued to receive them.

COMPUTER

2

Computer mode inhibits echoing of typed characters and is used when the port is
communicating with a computer operated control program.

HESSEN

PROTOCOL

16

The Hessen communications protocol is used in some European countries. TAI P/N
02252 contains more information on this protocol.

E, 8, 1

8192

When turned on this mode switches the COMM port settings from

NO PARITY; 8 data bits; 1 stop bit to EVEN PARITY; 8 data bits; 1 stop bit.

E, 7, 1

2048

When turned on this mode switches the COM port settings from

NO PARITY; 8 DATA BITS; 1 stop bit to EVEN PARITY; 7 DATA BITS; 1 stop bit.

RS-485

1024

Configures the COM2 Port for RS-485 communication. RS-485 mode has precedence
over multidrop mode if both are enabled.

SECURITY

4

When enabled, the serial port requires a password before it will respond (see Section
5.5). If not logged on, the only active command is the '?' request for the help screen.

MULTIDROP

PROTOCOL

32

Multidrop protocol allows a multi-instrument configuration on a single communications
channel. Multidrop requires the use of instrument IDs.

ENABLE

MODEM

64

Enables to send a modem initialization string at power-up. Asserts certain lines in the
RS-232 port to enable the modem to communicate.

ERROR

CHECKING

2

128

Fixes certain types of parity errors at certain Hessen protocol installations.

XON/XOFF

HANDSHAKE

2

256

Disables XON/XOFF data flow control also known as software handshaking.

HARDWARE

HANDSHAKE

8

Enables CTS/RTS style hardwired transmission handshaking. This style of data
transmission handshaking is commonly used with modems or terminal emulation
protocols as well as by Teledyne Instrument’s APICOM software.

HARDWARE

FIFO

2

512

Disables the HARDWARE FIFO (First In – First Out). When FIFO is enabled it improves
data transfer rate for that COM port.

COMMAND

PROMPT

4096

Enables a command prompt when in terminal mode.

1

Modes are listed in the order in which they appear in the

SETUPMORECOMMCOM[1 OR 2] MODE menu

2

The default setting for this feature is ON. Do not disable unless instructed to by TAI’s Customer Service personnel.

Communication Modes for each COM port must be configured independently. To turn
on or off the communication modes for either COM1 or COM2, access the
SETUP>MORE>[COM1 or COM2] menu and at the COM1[2] Mode menu press EDIT.