CommunicationsTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

149

5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following

settings:

Baud Rate: 115200
COM2 Mode Settings:
Quiet Mode

ON

Computer Mode

ON

MODBUS RTU OFF
MODBUS ASCII

OFF

E,8,1 MODE OFF
E,7,1 MODE OFF
RS-485 MODE OFF
SECURITY MODE

OFF

MULTIDROP MODE OFF
ENABLE MODEM

OFF

ERROR CHECKING ON
XON/XOFF HANDSHAKE

OFF

HARDWARE HANDSHAKE OFF
HARDWARE FIFO

ON

COMMAND PROMPT OFF

6. Next, configure your communications software, such as APIcom. Use the COM

port determined in Step 4 and the baud rate set in Step 5. The figures below show
how these parameters would be configured in the Instrument Properties window in
APIcom when configuring a new instrument. See the APIcom manual (PN 07463)
for more details.