Disabling/enabling data channels – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 179

Advertising
background image

DAC and APICONTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

179

7.1.9.2. THE STARTING DATE FEATURE

This option allows the user to specify a starting date for any given channel in case the
user wants to start data acquisition only after a certain time and date. If the STARTING
DATE
is in the past (the default condition), the DAS ignores this setting and begins
recording data as defined by the REPORT PERIOD setting.
The STARTING DATE generally cannot be accessed from the standard DAS front
panel menu, but is available via the instrument’s communication ports by using
APICOM or the analyzer’s standard serial data protocol.

7.1.10. DISABLING/ENABLING DATA CHANNELS

Data channels can be temporarily disabled, which can reduce the read/write wear on the
Disk-on-Module.
To disable a data channel, follow the instruction shown in Section 7.1.4 then press:

Advertising