2.2. EPA EQUIVALENCY DESIGNATION

Note

GFC7001TM: EPA equivalency does not apply to this model.

TAI’s Model GFC7001T, Carbon Monoxide Analyzer, is designated as Reference
Method Number RFCA-1093-093 as defined in 40 CFR Part 53, when operated under
the following conditions:

 Range: Any range from 10 ppm to 50 ppm
 Ambient temperature range of 10 to 40C
 Sample filter: Equipped with 5-micron PTFE filter element in the internal

filter assembly

 Software settings:

Dilution factor

1.0

AutoCal

ON or OFF

Dynamic Zero

ON or OFF

Dynamic Span

OFF

Dual Range

ON or OFF

Auto Range

ON or OFF

Temp/Pres Compensation

ON

Under the designation, the analyzer may be operated with or without the following
options:
 Rack mount with slides
 Rack mount without slides, ears only
 Zero/span valve options

 Option 50A – Sample/Cal valves, or
 Option 50B – Sample/Cal valves with span shutoff & flow control

 Internal zero/span (IZS) option with either:

 Option 51A – Sample/Cal valves, or
 Option 51C – Sample/Cal valves with span shutoff & flow control

 4-20mA, isolated output

2.3. APPROVALS AND CERTIFICATIONS

The TAI Model GFC7001T/GFC7001TM analyzer was tested and certified for Safety
and Electromagnetic Compatibility (EMC). This section presents the compliance
statements for those requirements and directives.

Accuracy

<± (0.02% CO

2

+ 2% of reading)

Linearity

<± 0.1% CO

2

Temperature Coefficient

<± 0.01% CO

2

/°C

Rise and Fall Time

<60 seconds to 95%

1

As defined by the USEPA

2

Defined as twice the zero noise level by the USEPA