Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 166

Advertising
background image

DAC and APICONTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

166

These default Data Channels can be used as they are, or they can be customized from the
front panel to fit a specific application. They can also be deleted to make room for
custom user-programmed Data Channels.
Appendix A-5 lists the firmware-specific DAS configuration in plain-text format. This
text file can either be loaded into APICOM and then modified and uploaded to the
instrument or can be copied and pasted into a terminal program to be sent to the
analyzer.

I

MPORTANT

I

MPACT ON

R

EADINGS OR

D

ATA

Sending a DAS configuration to the analyzer through its COM ports will
replace the existing configuration and will delete all stored data. Back up
any existing data and the DAS configuration before uploading new
settings.